10 SENSe Subsystem
408 Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:V2P ATYPe|DTYPe
This command is used to select the type of linearity correction that is applied to
the channel sensors being used. For most 8480 Series sensors the correct (A type
or D type) linearity correction table is automatically selected. However, for the
V8486A and W8486A sensors the automatic selection must be overridden and the
D type (diode) correction selected.
Syntax
Example
Reset Condition
On reset, the linearity correction is set for A type.
Query
[SENSe[1]]|SENSe2:V2P?
The query returns the current type of linearity correction being used (A type or
D type).
SENS2:V2P DTYP This command selects the D type linearity
correction to be applied to Channel B.