11 STATus Subsystem
428 Keysight N8262A Programming Guide
STATus:DEVice:NTRansition is 1, then STATus:DEVice:EVENt? is set
when a power sensor is disconnected.
STATus:DEVice:PTRansition is 1, then STATus:DEVice:EVENt? is set
when a power sensor is connected.
The Channel A and B sensor error bits (3 and 4) are set to:
1, if the P-Series or E-Series power sensor EEPROM has failed.
0, for every other condition.
The Soft front panel button click bit (bit 14), when queried with the
STATus:DEVice:EVENt? query indicates whether any soft front panel softkeys
have been clicked since power up or since you last queried the device status
register. This bit ignores the :NTRansition, and :PTRansition registers and a
:CONDition? query always returns a 0.
NOTE
Querying STATus:DEVice:EVENt? clears the STATus:DEVice:EVENt? register.