SYSTem Subsystem 12
Keysight N8262A Programming Guide 521
TD-SCDMA
The following table shows the power meter presets when <character_data> is
set to TDSCdma. Commands not listed are preset according to their DEFault values
(for further information refer to Table 12-1).
Table 12-37 1xeV-DV: Power meter presets for secondary channel sensors
Function Secondary Channel Sensor
No Sensor Non P-Series Sensor P-Series Sensor
Display setup
Lower window Dual numeric Dual numeric Dual numeric
Lower window/lower measurement (LL)
Feed
Gate 1 primary channel
[a]
Secondary channel
[a]
Gate 1 secondary channel
[a]
(Channel B)
Measurement Pk-to-Avg Avg Avg
[a] For further information refer to “Primary and Secondary Channels” on page 473.
Table 12-38 TD-SCDMA: Power meter presets
Command Setting Comments
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX] +1900.000 MHz Frequency setting
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC NORM Measurement mode
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID N1921/2A: LOW Sensor video bandwidth
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4:OFF:TIME Gate 1: 10 µs
Gates 2 - 4: 0
Delay between trigger point and time
gated period.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals