TRACe Subsystem 13
Keysight N8262A Programming Guide 539
:DURation? [query only] page 554
:OCCurrence? [query only] page 555
:POSitive
:DURation? [query only] page 556
:OCCurrence? [query only] page 557
:REFerence? <numeric_value> [query only] page 558
:STATe <boolean> page 560
:UNIT <character_data> page 562
Keyword Parameter Form Notes Page
When making trace measurements, use the following command sequence to
synchronize the returned trace data with the measurement:
Command Comment
INIT:CONT OFF Trace data can only be retrieved with INIT:CONT OFF
TRAC:STAT ON Enables trace capture
AVER:STAT OFF
or
TRIG:DEL:AUTO OFF
No settling time delays for digital filter to fill
INIT Initiates a new measurement
FETCH? Fetch the result (waits for the measurement to complete)
TRACE:DATA? MRES Retrieves the trace data once the measurement has completed