TRIGger Subsystem 14
Keysight N8262A Programming Guide 589
Example
Reset Condition
On reset the trigger holdoff is set to 1 µs.
Query
TRIGger[:SEQuence]:HOLDoff?
The query returns the current trigger holdoff setting.
Query Example
TRIG:SEQ1:HOLD 0.1 This command sets the trigger holdoff to 100
ms for Channel A.
TRIG:SEQ:HOLD? This command queries the trigger holdoff
setting for Channel A.