1 Power Meter Remote Operation
76 Keysight N8262A Programming Guide
Example
The following example sets the expected power value to –50 dBm and the
resolution setting to 3 using the lower level commands. The measurement is a
single Channel A measurement carried out on the lower window.
ABOR1 Aborts Channel A
CALC2:MATH:EXPR "(SENS1)" Displays Channel A on lower window
SENS1:POW:AC:RANG -50 Sets lower range (E-Series sensors only)
DISP:WIND2:RES 3 Sets the lower window’s resolution to setting 3
INIT1 Causes Channel A to make a measurement
FETC2? Retrieves the lower window’s measurement