Jitter analysis and clock data recovery
R&S
®
RTP
1035User Manual 1337.9952.02 ─ 12
Statistics
Enables the calculation and display of statistical results. They include:
"Min (abs/pp)": absolute minimum value of the component.
"Max (abs /pp)": absolute maximum value of the component.
"Min (σ)": minimum value of the standard deviation.
"Max (σ)": maximum value of the standard deviation.
"Wave count": Number of waveforms (acquisitions) the measurement is based on.
In the "Advanced Jitter Results" table, you can reset the statistics.
Remote command:
ADVJitter<m>:RESult:STATistics on page 2271
Step response
Enables a step response diagram. The diagram displays the time behavior of the sys-
tem when the inputs change from zero to one or from one to zero within a short time
period.
You can set the Length and the Step position for the step response diagram in the
"Advanced Settings" dialog.
For further analysis of the step response, you can use it as a source for "Amp/Time"
measurements. See Chapter 8.2.5, "Amplitude/time measurements", on page 351.
Remote command:
ADVJitter<m>:RESult:STEPresponse on page 2272
Component specific results
For the selected jitter component, selected you can display:
"Histogram": displays a histogram of measurement results for the selected compo-
nent - the cumulative occurrence distribution of measurement results in a graphic.
You can enable a cursor and use it for further analysis.
Advanced jitter and noise (option R&S
RTP- K133/K134)