Remote control commands
R&S
®
RTP
1193User Manual 1337.9952.02 ─ 12
Parameters:
<RecordLength> Number of recorded waveform points.
Range: 1000 to 1000000000
Increment: 2
*RST: 1000
Default unit: pts
Usage: Asynchronous command
TIMebase:ROLL:ENABle <Mode>
Activates the automatic roll mode.
Parameters:
<Mode> AUTO | OFF
AUTO: the instrument activates the roll mode under specific con-
ditions.
See also "Roll mode" on page 143.
*RST: AUTO
Usage: Asynchronous command
TIMebase:ROLL:STATe?
Returns the status of the roll mode.
Return values:
<State> ON | OFF
*RST: OFF
Usage: Query only
Asynchronous command
TIMebase:ROLL:MTIMe <MinHorizGain>
The roll mode is enabled automatically if the acquisition time exceeds the given value,
and if TIMebase:ROLL:ENABle is set to AUTO.
Parameters:
<MinHorizGain> Treshold value for roll mode enabling.
Range: 1 to 600
Increment: 1
*RST: 10
Default unit: s
Usage: Asynchronous command