Remote control commands
R&S
®
RTP
1193User Manual 1337.9952.02 ─ 12
Parameters:
<RecordLength> Number of recorded waveform points.
Range:  1000  to  1000000000
Increment:  2
*RST:  1000
Default unit: pts
Usage:  Asynchronous command
TIMebase:ROLL:ENABle <Mode>
Activates the automatic roll mode.
Parameters:
<Mode> AUTO | OFF
AUTO: the instrument activates the roll mode under specific con-
ditions.
See also "Roll mode" on page 143.
*RST:  AUTO
Usage:  Asynchronous command
TIMebase:ROLL:STATe?
Returns the status of the roll mode.
Return values: 
<State> ON | OFF
*RST:  OFF
Usage:  Query only
Asynchronous command
TIMebase:ROLL:MTIMe <MinHorizGain>
The roll mode is enabled automatically if the acquisition time exceeds the given value,
and if TIMebase:ROLL:ENABle is set to AUTO.
Parameters:
<MinHorizGain> Treshold value for roll mode enabling.
Range:  1  to  600
Increment:  1
*RST:  10
Default unit: s
Usage:  Asynchronous command
Acquisition and setup