Remote control commands
R&S
®
RTP
1197User Manual 1337.9952.02 ─ 12
●
It defines the number of waveforms used to calculate the average waveform.
Parameters:
<MaxAcqCnt> Range: 1 to 16777215
Increment: 10
*RST: 1
Usage: Asynchronous command
ACQuire:ARESet:IMMediate
Forces the immediate restart of the envelope and average calculation for all wave-
forms.
Usage:
Event
Asynchronous command
ACQuire:ARESet:MODE <ArtmRst>
Defines when the envelope and average evaluation restarts.
Parameters:
<ArtmRst> NONE | TIME | WFMS
TIME
Restarts the envelope and average calculation after the time
defined with ACQuire:ARESet:TIME.
WFMS
Restarts the envelope and average calculation after a number of
acquired waveforms defined with ACQuire:ARESet:COUNt
on page 1198.
*RST: NONE
Usage: Asynchronous command
ACQuire:ARESet:TIME <EnvelopeTimeout>
Defines the time after which the envelope and average evaluation restarts.
The setting is relevant if ACQuire:ARESet:MODE is set to TIME.
Parameters:
<EnvelopeTimeout> Range: 0.1 to 10000
Increment: 0.01
*RST: 0.1
Default unit: s
Usage: Asynchronous command
Acquisition and setup