Remote control commands
R&S
®
RTP
1279User Manual 1337.9952.02 ─ 12
<n> 1..31
Indicates the trigger source:
1...4 = channel 1 to 4
5 = external trigger input
6...11 = not available
12...27 = R&S RT-ZVC input channels
28...31 = differential signals Diff1, Diff2 and Common1, Com-
mon2
Parameters:
<Level> Voltage for the trigger level.
Range:  Depends on vertical scale, channel offset and other
settings. The trigger level must be within the current
display range.
Increment:  1E-3
*RST:  0
Default unit: V
Example: 
TRIG:LEV5 0.01
Sets the trigger level for the external trigger signal to 10 mV.
TRIG2:LEV3 0.2
Sets the trigger level for the B-event and B-trigger source chan-
nel 3 to 200 mV.
Usage:  Asynchronous command
TRIGger<m>:SPATtern:CSOurce[:VALue] <ClockSource>
Defines the source of the clock signal.
Suffix: 
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockSource> CHAN1 | CHANNEL1 | CHAN2 | CHANNEL2 | CHAN3 |
CHANNEL3 | CHAN4 | CHANNEL4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
Input channel of the clock signal
*RST:  CHAN1
Usage:  Asynchronous command
TRIGger<m>:SPATtern:CSOurce:EDGE <ClockEdge>
Together with the clock level (see TRIGger<m>:SPATtern:CSOurce:LEVel
on page 1280), the clock edge defines the point in time when the state of the data sig-
nal is checked.
Trigger