Remote control commands
R&S
®
RTP
1375User Manual 1337.9952.02 ─ 12
Usage: Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:DIRection <EdgeCntDirct>
Selects the direction for counting slopes for each source: from the beginning of the
waveform, or from the end.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
<n> 1..2
Selects the source number.
Parameters:
<EdgeCntDirct> FRFI | FRLA
FRFI - FRom FIrst, counting starts with the first edge of the
waveform.
FRLA - FRom LAst, counting starts with the last edge of the
waveform.
*RST: FRFI
Usage: Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:ECOunt <EdgeIndex>
Sets the number of the edge that is relevant for delay measurement for each source.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
<n> 1..2
Selects the source number.
Parameters:
<EdgeIndex> Edge number
Range: 1 to 100000
Increment: 1
*RST: 1
Usage: Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:LSELect <DlyLevSel>
Selects the reference level on which the time is measured for each source.
Automatic measurements