Remote control commands
R&S
®
RTP
1376User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
<n> 1..2
Selects the source number.
Parameters:
<DlyLevSel> UPPer | MIDDle | LOWer
*RST: MIDDle
Usage: Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:SLOPe
<Slope>
Sets the edge of each source, between which the delay is measured.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
<n> 1..2
Selects the source number.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
MEASurement<m>:AMPTime:PTCount <PulseCount>
Sets the number of positive pulses for the pulse train measurement. It measures the
duration of N positive pulses from the rising edge of the first pulse to the falling edge of
the N-th pulse.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<PulseCount> Range: 1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command