Remote control commands
R&S
®
RTP
1396User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<Number> Maximum number of peaks
Query parameters:
<MeasType> PLISt
Only for peaklist measurements. For other measurement, this
command returns an error.
MEASurement<m>:RESult:SHFRequency <MeasType>, <State>
MEASurement<m>:RESult:SHFRequency? <MeasType>
Includes the frequency of the detected peak in the diagram labels (if
MEASurement<m>:RESult:LABorder=ON).
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<State> ON | OFF
Parameters for setting and query:
<MeasType> PLISt
Only for peaklist measurements. For other measurements, this
command returns an error.
Example: Display labels.
MEAS:RES:SHL PLIS,ON
Show frequency in labels.
MEAS:RES:SHFR PLIS,ON
Query the type of labels for peak lists.
MEAS:RES:SHFR? PLIS
//Result: ON
MEASurement<m>:RESult:SHLabels <MeasType>, <State>
MEASurement<m>:RESult:SHLabels? <MeasType>
Defines whether a description (label) is displayed for each detected peak in the spec-
trum diagram.
The layout of the label is defined by MEASurement<m>:RESult:LABorder.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Automatic measurements