Remote control commands
R&S
®
RTP
1719User Manual 1337.9952.02 ─ 12
Usage: Query only
BUS<m>:ARINc:WORD<n>:PATTern?
Returns all 32 bits of the specified word.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Return values:
<Data> Comma-separated list of 4 bytes in big endian order. The format
of each byte is defined by FORMat:BPATtern.
Example:
BUS2:ARINc:WORD3:PATTern?
--> #H75,#H11,#H55,#H82
FORMat:BPATtern DEC
BUS2:ARINc:WORD3:PATTern?
--> 117,17,85,130
Usage: Query only
BUS<m>:ARINc:WORD<n>:SDI?
Returns the source/destination identifier (SDI) bits of the specified word.
Suffix:
<m>
.
1..4
<n> *
Return values:
<SDI> Range: 0 to 3
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:ARINc:WORD<n>:SSM?
Returns the sign/status matrix(SSM) bits of the specified word.
Suffix:
<m>
.
1..4
<n> *
Return values:
<SSM> Range: 0 to 3
Increment: 1
*RST: 0
Usage: Query only