Remote control commands
R&S
®
RTP
1800User Manual 1337.9952.02 ─ 12
Parameters for setting and query: 
<SearchName>
SEARch:TRIGger:TBTO:DADDress:MIN <SearchName>,<DestAddrPattMin>
SEARch:TRIGger:TBTO:DADDress:MIN? <SearchName>
Specifies the destination address bit pattern, or sets the start value of a pattern range.
Parameters:
<DestAddrPattMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
Parameters for setting and query: 
<SearchName>
SEARch:TRIGger:TBTO:DATA:DCONdition <SearchName>,<DataOperator>
SEARch:TRIGger:TBTO:DATA:DCONdition? <SearchName>
Sets the condition for the data. You can define an exact value or a value range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding SEARch:TRIGger:TBTO:
DATA:DMIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with SEARch:TRIGger:TBTO:DATA:DMIN and
SEARch:TRIGger:TBTO:DATA:DMAX.
*RST: 
EQUal
Parameters for setting and query: 
<SearchName>
SEARch:TRIGger:TBTO:DATA:DMAX <SearchName>,<DataMax>
SEARch:TRIGger:TBTO:DATA:DMAX? <SearchName>
Sets the end value of the data range if SEARch:TRIGger:TBTO:DATA:DCONdition
is set to INRange or OORange.
Parameters:
<DataMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
Protocols