Remote control commands
R&S
®
RTP
1822User Manual 1337.9952.02 ─ 12
TRIGger<m>:RFFE:DATA:ICONdition <DataIdxOperator>
Sets the operator to define the data frames in which the data pattern is expected.
Parameters:
<DataIdxOperator> EQUal | LTHan | LETHan | GTHan | GETHan | INRange |
RANGe
EQUal | LTHan | LETHan | GTHan | GETHan
Equal, Less than, Less or equal than, Greater Than, Greater or
equal than. These conditions require one pattern to be set using
TRIGger<m>:RFFE:DATA:IMIN.
INRange = RANGe
In range: Set the minimum and maximum value using
TRIGger<m>:RFFE:DATA:IMIN and TRIGger<m>:RFFE:
DATA:IMAX.
*RST: INRange
TRIGger<m>:RFFE:DATA:IMIN
<DataIndexMin>
Specifies a pattern, or sets the start value of the index range.
Parameters:
<DataIndexMin> Range: 1 to 32
Increment: 1
*RST: 1
TRIGger<m>:RFFE:DATA:IMAX <DataIndexMax>
Sets the end value of a range, if the index condition is set to INRange or OORange.
See TRIGger<m>:RFFE:DATA:ICONdition.
Parameters:
<DataIndexMax> Range: 1 to 32
Increment: 1
*RST: 32
TRIGger<m>:RFFE:SADD:CONDition <SIDOperator>
Sets the operator to trigger on a specific slave address/MID pattern or a range.
Parameters:
<SIDOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one pattern to
be set with TRIGger<m>:RFFE:SADD:MIN.
Protocols