Remote control commands
R&S
®
RTP
1824User Manual 1337.9952.02 ─ 12
Parameters:
<MaskMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:RFFE:MASK:MIN <MaskMin>
Specifies a mask pattern, or sets the start value of a mask pattern range.
Parameters:
<MaskMin>
TRIGger<m>:RFFE:INTerrupt
<Interrupt>
Sets the interrupt pattern for the interupt summary and notification trigger type.
Parameters:
<Interrupt>
TRIGger<m>:RFFE:ERRor:BP <BusParkError>
Enables the trigger on a bus park error - when an incorrect bus parlk cycle has been
found.
Parameters:
<BusParkError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:LENGth <LengthError>
Enables the trigger on a length error - when an incorrect length of the command
sequence has been found.
Parameters:
<LengthError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:NOResponse <NoResponse>
Enables the trigger on any No Response Frame.
Parameters:
<NoResponse> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:PARity <ParityError>
Enables the trigger on incorrect parity bits.
Protocols