Remote control commands
R&S
®
RTP
1825User Manual 1337.9952.02 ─ 12
Parameters:
<ParityError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:SSC <SSCError>
Enables the trigger on an SSC error - when no valid SSC sequence has been found
after the idle time.
Parameters:
<SSCError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:USEQuence
<UnknownSequence>
Enables the trigger on an unknown sequence - when the instrument cannot detect any
supported command sequence.
Parameters:
<UnknownSequence>ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:GAP <GapError>
Enables the trigger on a gap error.
Parameters:
<GapError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:ERRor:VERSion <VersionError>
Enables the trigger on RFFE version error.
Parameters:
<VersionError> ON | OFF
*RST: ON
TRIGger<m>:RFFE:INTerrupt <Interrupt>
Sets the interrupt pattern for the interupt summary and notification trigger type.
Parameters:
<Interrupt>