Remote control commands
R&S
®
RTP
1938User Manual 1337.9952.02 ─ 12
Parameters:
<SourceNRZEnable> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
BUS<m>:CMSB:NRZ:ENAPolarity <PolarityEnable>
Sets whether the transmitted enable signal is active when the voltage is below the
threshold (ENALow) or higher than it (ENAHIgh).
Suffix:
<m>
.
1..4
Parameters:
<PolarityEnable> ENALow | ENAHigh
*RST: ENAHigh
BUS<m>:CMSB:NRZ:POLarity <PolarityData>
Selects the polarity of the custom serial bus data signal in NRZ coding standards.
Suffix:
<m>
.
1..4
Parameters:
<PolarityData> ACTLow | ACTHigh
ACTLow
Active low: the value "1" is represented by a voltage below the
threshold
ACTHigh
Active high: the value "1" is represented by a voltage above the
threshold
*RST: ACTHigh
BUS<m>:CMSB:NRZ:THReshold:CLCK <ThresholdClock>
Sets the threshold for the clock signal digitization in the NRZ Clocked coding standard.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdClock> Range: -25 to 25
Increment: 0.1
*RST: 2
Default unit: V
BUS<m>:CMSB:NRZ:HYSTeresis:CLCK <HystClock>
Sets the hysteresis size on the clock channel.