Remote control commands
R&S
®
RTP
2093User Manual 1337.9952.02 ─ 12
Usage: Query only
BUS<m>:USBPd:RESult:FRAMe<n>:INFO?
Returns information on the frame with the selected frame number (for the selected
serial bus).
Suffix:
<m>
.
1..4
<n> *
Return values:
<Info>
Usage: Query only
BUS<m>:USBPd:RESult:FRAMe<n>:STARt?
Returns the start time of the frame with the selected frame number (for the selected
serial bus).
Suffix:
<m>
.
1..4
<n> *
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:USBPd:RESult:FRAMe<n>:STATe?
Returns the status of the frame with the selected frame number (for the selected serial
bus).
Suffix:
<m>
.
1..4
<n> *
Return values:
<State> OK | UNKNown | INComplete | PREamble | SOP | CRC |
LENGth | FBFB
SOP
Start of packet warning
CRC
Cyclic redundancy check error
Protocols