Remote control commands
R&S
®
RTP
2096User Manual 1337.9952.02 ─ 12
Parameters for setting and query:
<SearchName>
Example:
SEARch:TRIGger:USBPd:ERRor1:ENABle ON
Enables searching for CRC Error.
SEARch:TRIGger:USBPd:ERENable "Length Error",ON
Enables searching for Length Error.
SEARch:TRIGger:USBPd:FRENable <SearchName>,<Frame>, <Enabler>
SEARch:TRIGger:USBPd:FRAMe<m>:ENABle <SearchName>,<Enable>
SEARch:TRIGger:USBPd:FRAMe<m>:ENABle? <SearchName>
Suffix:
<m>
.
*
Specifies the frame number.
Parameters:
<Enable> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
Example:
SEARch:TRIGger:USBPd:FRAMe1:ENABle ON
Enables the checking condition for the Test frame.
SEARch:TRIGger:USBPd:FRENable "Control Frame"
ON
Enables the checking condition for the Control frame.
SEARch:TRIGger:USBPd:FIENable <SearchName>,<Frame>, <Field>, <Enabler>
SEARch:TRIGger:USBPd:FRAMe<m>:FLD<n>:ENABle
<SearchName>,<CondEnabler>
SEARch:TRIGger:USBPd:FRAMe<m>:FLD<n>:ENABle? <SearchName>
Enables or disables the checking condition for searching a specific data pattern in the
selected field of the selected frame.
Suffix:
<m>
.
*
Specifies the frame.
<n> *
Specifies the field number within the frame.
Parameters:
<CondEnabler> ON | OFF
ON
Checking condition enabled
OFF
Checking condition disabled
*RST: OFF
Protocols