Remote control commands
R&S
®
RTP
2179User Manual 1337.9952.02 ─ 12
Parameters for setting and query:
<SearchName> The following frames are available:
Read
Write
Read NC
Write NC
SEARch:TRIGger:DDR:IMAX <SearchName>,<Frame>, <Field>, <Data>
SEARch:TRIGger:DDR:FRAMe<m>:FLD<n>:IMAX <SearchName>,<IndexMax>
SEARch:TRIGger:DDR:FRAMe<m>:FLD<n>:IMAX? <SearchName>
Sets the end value of the bit index range (data position range), if SEARch:TRIGger:
DDR:FRAMe<m>:FLD<n>:IOPerator is set to INRange or RANGe.
Suffix:
<m>
.
*
<n> *
Parameters:
<IndexMax> Range: 1 to 65534
Increment: 1
*RST: 65534
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:DDR:IMIN <SearchName>,<Frame>, <Field>, <Data>
SEARch:TRIGger:DDR:FRAMe<m>:FLD<n>:IMIN <SearchName>,<IndexMin>
SEARch:TRIGger:DDR:FRAMe<m>:FLD<n>:IMIN? <SearchName>
Sets the index at which the data is to be searched, or sets the start value of an index
range in which the data is to be searched.
Suffix:
<m>
.
*
<n> *
Parameters:
<IndexMin> Specifies the index value or the start value of an index range
within the field.
The index range, increment and *RST values depend on the
field type.
Range: 1 to 65534
Increment: 1
*RST: 1
Parameters for setting and query:
<SearchName>