Remote control commands
R&S
®
RTP
2301User Manual 1337.9952.02 ─ 12
DEEMbedding<m>:COMPonent<n>:OPONe.................................................................. 2301
DEEMbedding<m>:COMPonent<n>:OPTWo.................................................................. 2301
DEEMbedding<m>:COMPonent<n>:SPONe...................................................................2302
DEEMbedding<m>:COMPonent<n>:SPTWo...................................................................2302
DEEMbedding<m>:COMPonent<n>:IDENtical <MatchedPaths>
The setting is relevant if the measurement setup is a 4-port, and the component is a 2-
port. Usually, these components are cables and adapters, which are described by
*.s2p files. In case of a 4-port setup, 2 cables or 2 adapters are used. If they are identi-
cal, for example, when a matched-pair cable is used, they can be described by the
same file. In this case, set the command to ON.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1..*
Selects the component. The DUT has suffix 1.
Parameters:
<MatchedPaths> ON | OFF
*RST: OFF
Usage: Asynchronous command
DEEMbedding<m>:COMPonent<n>:IPONe <InputPort1>
DEEMbedding<m>:COMPonent<n>:IPTWo <InputPort2>
DEEMbedding<m>:COMPonent<n>:OPONe <OutputPort1>
DEEMbedding<m>:COMPonent<n>:OPTWo <OutputPort2>
Assign the ports to the input and output of the component according to the data in the
Touchstone file. This information is not directly written in the file. You need background
information how the S-parameters were determined, i.e. which port was input and
which port was output.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
<n> 1..*
Selects the component. The DUT has suffix 1.
Parameters:
<InputPort1>
<InputPort2>
<OutputPort1>
<OutputPort2>
ONE | TWO | THRee | FOUR
Port number that is assigned to the input or output.
*RST: ONE for input 1, TWO for output 1, THRee for
input 2, FOUR for output 4
Usage: Asynchronous command
Deembedding (Option R&S RTP-K121/-K122)