Measurements
R&S
®
RTP
323User Manual 1337.9952.02 ─ 12
6. Select the cursor type: horizontal, vertical, or both.
7. Define the position of the cursors:
a) To define exact positions of the cursor lines, enter the X-position for each verti-
cal cursor and the Y-position for each horizontal cursor. If it is not possible to
set horizontal cursors, disable "Track waveform".
b) To position the horizontal cursors automatically, select "Track waveform".
In this case, cursor 1 indicates the current maximum, cursor 2 indicates the
current minimum. If both horizontal and vertical cursors are displayed, the hori-
zontal cursors are placed at the crossing points of the vertical cursors with the
waveform. Adjust the vertical cursors manually, and the horizontal cursors fol-
low.
c) If the waveform arithmetics are set to "Envelope", and "Track waveform" is
active, select which horizontal cursor is positioned to the maximum and which
to the minimum envelope values.
d) To keep the distance between the vertical cursors when one cursor is moved,
select "Coupling (2 follows 1)".
8. Optionally, select "Show in all diagrams" in the "Advanced" tab. This setting ena-
bles the cursor display in all diagrams that are in the same domain as the selected
source (time or spectrum).
9. To set the cursors for a spectrum measurement to peak values, select the "Peak
Search" tab.
Tap one of the buttons to place the cursors on the selected peak value. For details,
see Chapter 8.1.3.3, "Peak search tab", on page 328.
10. Tap the "Cursor On/Off" icon in the "Setup" tab to activate the cursor measurement.
The cursors lines and the results are displayed. For details on cursor measurement
results, see Chapter 8.1.1, "Cursors and results of cursor measurements",
on page 319.
8.1.2.3 Configuring the cursor display
By default, the cursors are displayed as lines in the diagrams and labeled according to
the syntax: Cu<cursor set number>.<1|2>
For example, the cursors for the cursor set 2 are labeled 2.1 and 2.2. The horizontal
and the vertical cursors lines have the same labels.
You can change the default cursor display and labels.
To set the cursor style
1. Open the "Menu" and select "Settings".
2. Select "Appearance".
3. Select the "Cursor" tab.
4. Select the cursor style. See also: "Cursor style" on page 107.
Cursor measurements