Measurements
R&S
®
RTP
353User Manual 1337.9952.02 ─ 12
Meas. type Symbol Description/Result
Amplitude X
Ampl
Amplitude of the signal: the difference of high and low signal lev-
els
X
Ampl
= X
High
- X
Low
Max X
Max
Absolute maximum value of the waveform
Min X
Min
Absolute minimum value of the waveform
Peak to peak X
PkPk
Peak-to-peak value of the waveform: the difference of maximum
and minimum values
X
Ampl
= X
Max
- X
Min
Mean X
Mean
Arithmetic average of the waveform voltage values
Eval
N
i
Eval
Mean
ix
N
X
1
)(
1
RMS X
RMS
RMS (root mean square, quadratic mean) of the waveform volt-
age values
Eval
N
i
Eval
RMS
ix
N
X
1
2
)(
1
σ
σ (S-dev)
σ
X
Standard deviation of the waveform samples
Pos. overshoot R
Pos
Positive overshoot of a square wave, calculated from measure-
ment values High, Max, and Amplitude
%100
Amp
Ptop
V
VV
Ovr
Neg. overshoot R
Neg
Negative overshoot of a square wave, calculated from measure-
ment values Min, Low, and Amplitude
%100
Amp
Pbase
V
VV
Ovr
Cycle mean
The mean value of one cycle
Cycle RMS
The RMS (root mean square) value of one cycle
σ
Cycle σ (S-dev)
The standard deviation of one cycle
Trig. ProbeMeter
The DC voltage from the connected probe. Only available if an
active Rohde & Schwarz probe with ProbeMeter is connected.
Automatic measurements