Measurements
R&S
®
RTP
391User Manual 1337.9952.02 ─ 12
Enabling the track enables also the Continuous auto scale and Measure all events in
each acquisition.
To analyze the track, you can use cursor measurements and zoom.
Creating a track waveform
1. Configure the measurement group.
2. Activate the measurement group.
3. Select the "Result Analysis" tab.
4. If several measurements are enabled, select the "Measurement for analysis /
math".
5. Enable the track.
See also: "Enable (Track)" on page 393.
8.2.11.5 Settings for result analysis
The settings in the "Result Analysis" tab activate and configure long-term measure-
ments, statistical calculations, and the histogram of measurement results over a longer
period of time.
For scaling settings of the long-term diagram, see Chapter 8.2.11.6, "Horizontal long-
term scaling", on page 394.