Measurements
R&S
®
RTP
392User Manual 1337.9952.02 ─ 12
Measurement for plot
Selects the measurement that is used as source for mathematic calculations, long-term
measurements, and histograms. These evaluations are performed on one measure-
ment only, not on all measurements of the group. Statistics are calculated for all mea-
surements.
The selected measurement cannot be disabled in the "Add / Remove Measurement"
dialog box.
Remote command:
MEASurement<m>:MAIN on page 1364
Enable (Long term)
Enables long-term measurement of the main measurement.
Long-term measurements are performed on the "Measurement for plot".
Remote command:
MEASurement<m>:LTMeas[:STATe] on page 1415
Enable (Histogram)
Displays a histogram of measurement results - the cumulative occurrence distribution
of measurement results in a graphic. Enabling the histogram enables also the statistics
and "Measure all events in each acquisition".
Long-term measurements are performed on the "Measurement for plot".
Automatic measurements