2 MEASurement Commands
136 Keysight N8262A Programming Guide
Example
resolution
A numeric value for the resolution. If unspecified the current
resolution setting is used.
1 to 4
[b]
1.0, 0.1, 0.01, 0.001
DEF
[a]
source list
The channel which the command is implemented on.
If unspecified the current window setup is used. However,
if the window shows a ratio or difference measurement, the
upper window defaults to Channel A and the lower window
to Channel B.
(@1)
(@2)
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
CONF2:REL -50DBM,3,(@1) This command configures the lower window/
upper measurement to measure the relative
power of Channel A, using an expected
power level of –50 dBm and a resolution
setting of 3.
Item Description/Default Range of Values