EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
MEASurement Commands 2
Keysight N8262A Programming Guide 149
Example
Error Messages
If the last measurement is not valid error –230, “Data corrupt or stale” occurs.
A measurement is valid after it has been initiated. It becomes invalid when
either a reset occurs or any measurement parameter, for example frequency, is
changed.
If the expected_value and resolution parameters are not the same as the
current expected value and resolution settings on the specified window, error
–221, “Settings conflict” occurs.
resolution A numeric value for the resolution. If it is unspecified the current
resolution setting is used. If a value is entered it should
correspond to the current resolution setting otherwise an error
occurs.
1 to 4
[b]
1.0, 0.1, 0.01, 0.001
DEF
[[a]
source list The channel which the command is implemented on. If
unspecified the current window setup is used. However, if the
window shows a ratio or difference measurement, the upper
window defaults to Channel A and the lower window to Channel B.
(@1)
(@2)
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
FETC1:REL? DEF,2,(@2)
This command queries the upper window/
upper measurement relative measurement of
Channel B, using the current sensor range
and a resolution setting of 2.
Item Description/Default Range of Values

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals