SENSe Subsystem 10
Keysight N8262A Programming Guide 373
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 Commands
These commands control the pulse power measurement feature of the power
meter.
The following commands are detailed in this section:
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut]
[:MAGNitude] <numeric_value>
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <boolean>
NOTE
You can use either DCYCLe or GAIN3 in these commands, both do the same.
Using GAIN3 complies with the SCPI standard whereas DCYCle does not, but
may make your program more understandable.