SENSe Subsystem 10
Keysight N8262A Programming Guide 381
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe <boolean>
This command is used to enable/disable a channel offset for the power meter
setup. The [SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]
command is used to enter the loss/gain value.
Syntax
Example
Reset Condition
On reset, channel offsets are disabled.
Query
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe?
The query enters 1 or 0 into the output buffer indicating the status of the channel
offsets.
– 1 is returned if a channel offset is enabled
– 0 is returned if a channel offset is disabled
CORR:GAIN2:STAT ON This command enables a channel offset on
Channel A.
SENS
1
:GAIN2
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?