10 SENSe Subsystem
400 Keysight N8262A Programming Guide
[SENSe[1]]|SENSe2:TEMPerature?
This this command to returns the P-Series power sensor's temperature in degrees
Celsius.
Syntax
Parameters
Example
Reset Condition
On reset, this parameter is not affected.
Error Messages
If a P-Series sensor is not connected, error –241, “Hardware missing” occurs.
SENS
1
:
TEMP
SENS2
:
?
Item Description/Default Range of Values
numeric_value A numeric value defining sensor's temperature in degrees
Celsius.
–50 to 100
SENS2:TEMP? This command returns the current sensor
temperature found on Channel B.