EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Power Meter Remote Operation 1
Keysight N8262A Programming Guide 71
Some fine tuning of measurements can be carried out using the CONFigure
command and INITiate and FETCh? commands. For example, in the above
program segment some fine tuning can be carried out by setting the filter length
to 1024 and the trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 INIT2
6 FETC1? -50,DEF,(@2)
Example 4 - Specifying the Resolution Parameter
The previous examples detailed the use of the expected value and source list
parameters. The resolution parameter is used to set the resolution of the specified
window. This parameter does not affect the resolution of the data, however it does
affect the auto averaging setting (refer to Figure 1-2 on page 85).
Since the filter length used for a channel with auto-averaging enabled is
dependent on the window resolution setting, a conflict arises when a given
channel is set up in both windows and the resolution settings are different. In this
case, the higher resolution setting is used to determine the filter length.
The following example uses the resolution parameter to specify a resolution
setting of 3. This setting represents 3 significant digits if the measurement suffix is
W or %, and 0.01 dB if the suffix is dB or dBm (for further details on the resolution
parameter refer to the commands in Chapter 2, "MEASurement Commands").
Also, in this example the expected power and source list parameters are
defaulted. The expected power value is left unchanged at its current setting. The
source list parameter is defaulted as described in the note “Keysight N8262Aon
page 65. Note that as the source list parameter is the last specified parameter you
do not have to specify DEF.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals