1 Power Meter Remote Operation
72 Keysight N8262A Programming Guide
Using READ?
Some fine tuning of the above program segment can be carried out for example,
by setting the trigger delay off. The following program segment assumes that
Channel A is currently being measured on the upper window.
1 ABOR1
2 CONF1 DEF,3
3 TRIG1:DEL:AUTO OFF
4 READ1?
Using INITiate and FETCh?
The following program segment assumes that Channel A is currently being
measured on the upper window.
ABOR1 Aborts Channel A
CONF1 DEF,3 Configures the upper window to make a measurement
using the current setting of the expected power and
source list and a resolution setting of 3
READ1? Takes the upper window’s measurement. This is Channel A
or B measurement depending on current window setup.
ABOR1 Aborts Channel A
CONF1 DEF,3 Configures the upper window to make a measurement
using the current setting of the expected power and
source list and a resolution setting of 3
INIT1 Causes Channel A to make a measurement
FETC1? DEF,3 Retrieves the upper window’s measurement