Power Meter Remote Operation 1
Keysight N8262A Programming Guide 89
10 !Create I/O path name
20 ASSIGN @POWER TO 713
30 !Clear the power meter’s interface
40 CLEAR @POWER
50 !Set the power meter to a known state
60 OUTPUT @POWER;"*RST"
70 !Configure the Power Meter to make the measurement
80 OUTPUT @Power;"CONF:POW:AC:RAT 20DBM,2,(@1),(@2)"
90 !Set the measurement units to dBm
100 OUTPUT @POWER;"UNIT:POW DBM"
110 !Set the power meter for channel offsets of -10 dB
120 OUTPUT @POWER;"SENS1:CORR:GAIN2 -10"
130 OUTPUT @POWER;"SENS2:CORR:GAIN2 -10"
140 !Enable the gain correction
150 OUTPUT @POWER;"SENS:CORR:GAIN2:STATe ON"
160 OUTPUT @POWER;"SENS2:CORR:GAIN2:STATe ON"
170 !Set the power meter for a display offset of -20 dB
180 OUTPUT @POWER;"CALC1:GAIN -20 DB"
190 PRINT "MAKING THE MEASUREMENT"
200 !Initiate the measurement
210 OUTPUT @Power;"INIT1:IMM"
220 OUTPUT @Power;"INIT2:IMM"
230 ! ... and get the result
240 OUTPUT @Power;"FETC:POW:AC:RAT? 20DBM,2,(@1),(@2)"
250 ENTER @Power;Reading
260 !
270 PRINT "The measurement result is ";Reading;"dB."
280 END
For further information on channel offsets refer to page 379. For further
information on display offsets refer to page 192.
A
dBm
10–
B
dBm
10–
----------------------



20–



dB

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals