4. MEASUREMENT FLOW
3 Confirm that the stylus is in proper contact with the measured surface.
In addition, confirm that the detector is parallel to the measured surface.
• Front view of detector
90°
Detector
Measured
surface
If the detector is not parallel to the measured surface, loosen the screw (M4) securing the
drive unit to the height/tilt adjustment unit. Adjust the detector so that it is parallel to the
measured surface, then secure the drive unit again.
• Side view of detector (skidless)
If the detector is not parallel to the measured surface, turn the height/tilt adjustment on the
height-tilt adjustment unit to adjust the detector.
• Side view of detector (skid-attached)
If the detector is not parallel to the measured surface, turn the height/tilt adjustment on the
height-tilt adjustment unit to adjust the detector.
4 When the measurement is started from an arbitrary position, adjust the
measurement start position by moving the drive unit from Drive Unit Operation
screen.
■ Measurable level meter
When the detector is attached to the drive/detector unit, it is possible to check whether or
not the detector is in the measurable position on the Home screen.
Measurable level meter
Detector
Measured
surface
Measured
surface
Detector
Home screen
No. 99MBB765A
4-13