4.5 Perform Measurement
To start measurement, set the drive/detector unit on a workpiece and press . While
measurement is being performed, the evaluation profile is displayed. After measurement
has been completed, the measurement result will be displayed for confirmation.
4.5.1 Set the workpiece and drive/detector unit.
Setting the workpiece and drive unit
For surface roughness measurement to be successful, it should be performed on a firm
base that is insulated as well as possible from all sources of vibration. If measurement is
performed being subject to significant vibrations, results may be unreliable.
TIP
In cases where the measured surface is smaller than the drive unit or where the surface
is curved (cylindrical, etc.), install the SJ-410 using an appropriate optional accessory
such as the simple column stand. For information about optional accessory, refer to
"22.8 Optional Accessories" (page 22-9).
1 Position the work piece so that the measured surface is level.
2 Set the drive/detector unit. Subsequently, perform vertical positioning and leveling to
adjust the height and angle of the drive/detector unit.
TIP
For information about the vertical positioning, refer to "4.2.1 Perform the vertical
positioning
" (page 4-2).
For information about the leveling, refer to "
4.2.4 Perform leveling" (page 4-4).
NOTE
For information about the short approach length measurement, refer to "5.3 Shortening
the Approach Length for Measurement" (page 5-6).
No. 99MBB765A