Loading...
21. TROUBLESHOOTING
No. 99MBB765A
21-5
21.3 Calculation Results
■ Calculation results
Symptom/Error
display
Possible causes
Remedies
Abnormal calibration
value!
The result of calibration measurement exceeds
the possible range of calibration.
Check the value of the precision roughness
specimen and the inputted nominal value.
Also, check the calibration measurement setup
conditions.
L3.000 um (118.110 uin)
Measurement result under a condition of
insufficient number of peaks and valleys.
C3,000 um
Measurement result under a condition that data
deletion is specified.
V3.000 um (118.110 uin)
Measurement result under a condition that
over-range has occurred.
S3.000 um (118.110 uin)
Measurement result obtained by calculation after
it is aborted.
Err014
Conditions for calculating a mean line in tilt
compensation were not appropriate.
Err110
Parameter can not be calculated due to
insufficient number of peaks and valleys.
Err115
Calculation cannot be performed due to
insufficient peaks and valleys.
E rr116
Equivalent line can not be calculated.
E rr117
Roughness motif can not be calculated as more
than 2 local peaks which have required height
does not exist.
E rr118
The first roughness motif exceeds upper limit of
the length A.
E rr119
Waviness motif can not be calculated as more
than 2 local peaks which have required height do
not exist.
E rr120
The first waviness motif exceeds upper limit of the
length B.
E rr121
Parameter can not be calculated as more than 3
motifs do not exist.

Table of Contents