No. 99MBB765A
21-6
Symptom/Error
display
Possible causes
Remedies
Calculation result is
abnormal.
Value is large/Value is
small/Value remain the
same irrespective of
workpiece
Detector is not properly connected to the drive
unit.
Properly connect the detector to the drive unit.
Refer to "3.3.1 Attaching and detaching the detector"
(page 3-6).
Connecting cable between drive unit and display
unit is not properly connected.
Properly connect the drive unit to the display unit.
Refer to "3.3.2 Plugging/unplugging the connection
cable" (page 3-12).
Calibration measurement was not performed
correctly due to improper setup of the SJ-410.
Re-calibrate the SJ-410.
Refer to "Chapter 6 CALIBRATION" (page 6-1).
Stylus is worn.
Or other than the above.
Contact your dealer or the nearest Mitutoyo sales
office.
The GO/NG Judgment
result indicator will not
appear.
The GO/NG Judgment parameter has not been
set.
Select desired parameter for setting the GO/NG
Judgment.
Refer to "8.3.7 Setting the GO/NG judgment function"
(page 8-20).
Upper limit/lower limit is set to minimum.
Set the upper limit or lower limit.
Refer to "8.3.7 Setting the GO/NG judgment function"
(page 8-20).
Calculation error
008
Number of files was insufficient for conic
correction process.
Calculation error
009
Appropriate profile was not applied in conic
correction.
Calculation error
011
Number of files was insufficient for calculation
process.
Calculation error
012
The range of tilt compensation is not correct.
Calculation error
013
Calculation cannot be performed due to excess
number of notched workpiece files.
Calculation error
014
Conditions for calculating a mean line in tilt
compensation were not appropriate.
Calculation error
015
Arm correction cannot be calculated.
Calculation error
016
A malfunction has occurred in the constant pitch
processing for arm correction calculation.
Calculation error
017
Number of files was insufficient for arm correction
calculation.
Calculation error
018
The conditions for tilt compensation are not
correct.
Calculation error
033
Excess number of maximum measurement data
files
Calculation error
034
Cutoff length (λc) setting error