Mitutoyo SURFTEST SJ-410 - Calculation Conditions When a Profile Motif (R-Motif;W-Motif) Is Selected; Setting a Motif Connection Method

Mitutoyo SURFTEST SJ-410
603 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
No. 99MBB765A
8-18
8.3.6 Calculation conditions when a profile motif (R-Motif/W-Motif) is selected
8.3.6.1 Setting a motif connection method
The SJ-410 can use either ISO 12085 Main Body compatible method or ISO 12085 Annex
A compatible method for the motif connection method when selecting a profile motif
(AnnexA).
■ Operating procedure (Refer to 8.1 "■ Accessing the Parameter Setup screen".)
Home screen to Main Menu
Details Setting screen
1
Touch .
Touching
cycles through the available settings,
ON and OFF.
AnnexA "ON" : Compliant with AnnexA
AnnexA "OFF": Compliant with main body of standard
Details Setting screen