7.3.2.5 Modifying the upper limit length (A) of R-Motif
When R-Motif is selected, the upper limit length (A) and the cut-off length (λs) are
displayed on the Evaluating Condition Setup screen as evaluation-related variable
parameters. Change upper limit length (A) and cut-off length (λs) if required.
■ Relationship between upper limit length (A) and cut-off length (λs)
When an upper limit length (A) is set, a cut-off length (λs) is set to one of the values as
shown below.
Upper limit length (A) Cut-off length (λs) Evaluation length (arbitrary length)
setting range
0.02 mm (0.001 in) 2.5 μm, 8 μm, 25 μm, (98.425 μin,
314.960 μin, 984.251 μin), NONE
0.10 - 0.64 mm (0.003 - 0.025 in)
0.1mm (0.003 in) 2.5 μm, 8 μm, 25 μm, (98.425 μin,
314.960 μin, 984.251 μin), NONE
0.65 - 3.20 mm (0.025 - 0.125 in)
0.5 mm (0.019 in) 2.5 μm, 8 μm, 25 μm, (98.425 μin,
314.960 μin, 984.251 μin), NONE
3.21 - 16.00 mm (0.126 - 0.629 in)
2.5 mm (0.098 in) 2.5 μm, 8 μm, 25 μm, (98.425 μin,
314.960 μin, 984.251 μin), NONE
16.01 - 25.40 (50.80) mm
0.063 - 1 (2) in
L (0.01 - 25.39 (50.79)mm)
(0.001 - 0.999 (1.999) in)
2.5 μm, 8 μm, 25 μm, (98.425 μin,
314.960 μin, 984.251 μin), NONE
0.10 - 25.40 (50.80) mm
0.003 - 1.0 (2.0) in
NOTE
• The evaluation length is saved for each upper limit length (A) item. Therefore, if the upper
limit length (A) is changed, the system will set the evaluation length that has been
previously stored at the changed upper limit length (A).
Relationship between evaluation length and upper limit length (A) must meet the condition
below:
Evaluation length > Upper limit length (A)
TIP
• For information on modifying the evaluation length, refer to "7.3.6 Setting the evaluation
length to an arbitrary length" (page 7-27).
No. 99MBB765A