23. REFERENCE INFORMATION
No. 99MBB765A
23-3
23.1.3 Evaluating based on VDA
Shown below are the standard cut-off values, sampling lengths, and evaluation lengths for
evaluation based on VDA.
NOTE
With SJ-410, when the VDA standard is selected, the λs filter automatically changes to
(NONE). To enable the λs filter, refer to "7.3.2 Modifying the Evaluation Profile and
Cut-off Length" (page 7-7).
Be aware that with the VDA standard, there are some differences with JIS B0601-2001
and ISO, such as λs not being set by default.
Standard sampling lengths and evaluation lengths for the measurement of Ra and Rq from non-periodic
roughness profiles
Ra Range
Sampling length (ℓ)
Evaluation length (ℓn)
0.006
< Ra 0.02 µm 0.08 mm 0.4 mm
0.02 < Ra 0.1 µm 0.25 mm 1.25 mm
0.1 < Ra 2.0 µm 0.8 mm 4 mm
2.0 < Ra 10.0 µm 2.5 mm 12.5 mm
10.0 < Ra 80.0 µm 8 mm 40 mm
Standard sampling lengths and evaluation lengths for measurement of Rz, Rp, and Rt from non-periodic
roughness profiles
Rz range
Sampling length (ℓ)
Evaluation length (ℓn)
0.025
< Rz 0.10 µm 0.08 mm 0.4 mm
0.10 < Rz 0.50 µm 0.25 mm 1.25 mm
0.50 < Rz 10.0 µm 0.8 mm 4 mm
10.0 < Rz 50.0 µm 2.5 mm 12.5 mm
50.0 < Rz 200.0 µm 8 mm 40 mm
Standard sampling lengths and evaluation lengths for the measurement of roughness parameters from
periodic profiles, and for the measurement of RSm from both periodic and non-periodic profiles
RSm Range
Sampling length (ℓ)
Evaluation length (ℓn)
0.013 < RSm 0.04 mm 0.08 mm 0.4 mm
0.04 < RSm 0.13 mm 0.25 mm 1.25 mm
0.13 < RSm 0.4 mm 0.8 mm 4 mm
0.4 < RSm 1.3 mm 2.5 mm 12.5 mm
1.3 < RSm 4.0 mm 8 mm 40 mm