No. 99MBB765A
6-14
Calibration Condition Setup screen
4
Touch .
Selected number of me
asurements will appear on the
Calibration Condition Setup screen.
6.4.3 Changing the calibration conditions according to the precision roughness
specimen
Set the following calibration conditions according to the precision roughness specimen:
Roughness standard, filter, cutoff length (λc)
NOTE
Be careful when modifying the calibration conditions, as the profile filter may be
automatically modified as a result.
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Screen".)
Home screen to Main Menu
Calibration Menu screen
1
Touch on the Calibration Menu screen.