6.4.4 Changing the calibration conditions according to the step specimen
Set the calibration conditions, such as traversing speed, returning speed, and range,
according to the step specimen.
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Screen".)
Home screen to Main Menu
Calibration Menu screen
1
Touch on the Calibration Menu screen.
Calibration Conditions screen
2
Touch .
Calibration Method Setup screen
3
Touch .