6. CALIBRATION
No. 99MBB765A
6-21
6.7 Executing Preliminary Measurement for
Calibration
To execute preliminary measurement, refer to "15 PRELIMINARY MEASUREMENT
(AUXILIARY FUNCTIONS)" (page 15-1).
6.8 Setting the Stylus Type
If the stylus is replaced with a stylus provided as an optional accessory to widen the
measuring range, the type of stylus set for this system must be changed.
TIP
If the stylus is detached and attached, or replaced, make sure to perform calibration
Without properly calibrating the instrument, correct measurements can not be obtained.
Various types of styli as well as the standard stylus are available for this system as
optional accessories.
The calibration values managed separately for each registered stylus. After you register
a stylus, be sure to perform calibration.
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Menu screen".)
Home screen to Main Menu
Calibration Menu screen
1
Touch on the Calibration Menu screen.
The Stylus Type Select window appears.
Stylus Select screen
2
Touch the stylus to be set.