5. OPERATION METHOD FOR THE DRIVE UNIT/DETECTOR/AUTOMATIC POSITIONING UNIT (OPTIONAL ACCESSORIES)
No. 99MBB765A
5-11
5.4.3 Switching the range
The SJ-410 can perform measurement in one of the following measuring ranges: 8, 80,
800, or AUTO.
The measurement resolution for each range is as shown in the following table.
Measurement range (µm)
800
80
8
Minimum resolution (µm)
0.0125
0.00125
0.000125
When the measurement range is set to AUTO, the measurement resolution varies
depending on the amount of displacement in the measurement range.
To perform a minute roughness evaluation, you can switch to further sensitive (narrow)
range for measurement.
TIP
To perform skidless measurement with further sensitive (narrow) range, you must adjust
the workpiece in the horizontal direction correctly. For information about adjustment in
the horizontal direction, refer to "4.2.4 Perform leveling" (page 4-4).
■ Operating procedure (Refer to 5.1 "■ Accessing the Drive Unit Operation screen".)
Home screen
Drive Unit Operation screen
1
Touch the measurement range button to modify to the target
measurement range.
Drive Unit Operation screen
The measurement range switches.