6. CALIBRATION
No. 99MBB765A
6-13
Calibration Measurement screen
5
Returns to the Calibration Measurement screen.
The Measurement screen for the selected calibration method
appears.
6.4.2 Setting the number of measurements
Specify the number of measurements for performing calibration.
The calibration results of specified numbers of measurements will be averaged to adjust
gain.
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Menu screen".)
Home screen to Main Menu
Calibration Menu screen
1
Touch on the Calibration Menu screen.
Calibration Condition Setup screen
2
Touch .
Number of Measurement Setup screen
3 Enter the number of measurement according to measurement
purpose.
TIP
Touching sets the value to 0.
For information about numeric value entry, refer to
"2.2.4 Entering numeric values/characters" (page 2-5).