21. TROUBLESHOOTING
No. 99MBB765A
21.2 Operation Error
■ Measuring operation
The result has exceeded the measurement range
during measurement.
- Adjust the detector stylus position.
- If it is tilted, adjust it in the preliminary
measurement.
- If the measurement range has been fixed, modify
the range to the larger one.
When the level meter displayed on the left of the
screen is red, over-range error occurs.
The limit terminal is being touched during
measurement.
Return the drive unit to the origin direction, and
perform the measurement again.
The detector is not properly connected.
Confirm that the detector is properly connected.
When the drive unit is properly connected and the
SJ-410 does not recover from the error, contact your
dealer or the nearest Mitutoyo sales office.
The drive unit is not properly connected.
Confirm that the drive unit is properly connected.
A current limit error has occurred while using the
automatic positioning unit.
Check the axis operation range for obstacles.
When this error occurs, you must restart the SJ-410.
is pressed during measurement.
Over-range has occurred.
Perform measurement again.
No measurement is
performed right after
is pressed.
Self-timer function has been set to ON.
Set the self-timer function to OFF.
Refer to "13.13.2 Setting the self-timer" (page 13-41).
Automatic positioning
function is not available
with the detector.
Automatic positioning does not function in the
detector depending on its type, such as a
skid-attached detector.
When printout is started, no printer paper is
loaded on the printer.
Load the printer paper on the printer.
Refer to "3.7 Setting the Printer Paper" (page 3-19).
The printer cover is not properly set.
Set the printer cover properly.
Refer to "3.7 Setting the Printer Paper" (page 3-19).