No. 99MBB765A
6-18
6.5 Confirming the Calibration History
SJ-410 can store up to 100 cases of date and time information as calibration history.
Confirm the calibration history following the procedures below.
NOTE
Be aware that the calibration history is completely cleared when [Clear all history]
is pressed.
Also, be aware that the calibration history is completely cleared, when power supply
from the built-in battery is cut or "RestToDefault" in the "Set Environ" is performed.
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Menu screen".)
Home screen to Main Menu
Calibration Menu screen
1
Touch on the Calibration Menu screen.
Calibration History screen
2
Confirm the data of time of the calibration.
TIP To cancel the calibration history, touch [Clear all
history], then
.

Table of Contents