6. CALIBRATION
No. 99MBB765A
6.6 Setting the Stylus Alarm
Stylus alarm is a function to notify when to replace the detector or perform regular
calibration by setting the threshold for cumulative measured value of the measuring length.
This section explains how to specify the threshold.
NOTE
• Be aware that the calibration history is completely cleared, when the battery switch is
set to OFF and power supply from built-in battery is cut, or "RestToDefault" in the "Set
Environ" is performed.
■ Operating procedure (Refer to 6.2 "■ Accessing the Calibration Menu screen".)
Home screen to Main Menu ⇒ ⇒
Calibration Menu screen
1 Touch on the Calibration Menu screen.
Stylus Alarm screen
Touch .
TIP
•
To clear the cumulative distance, touch
on
the Stylus Alarm screen, then
.
Threshold Setup screen
Specify the threshold.
TIP • Touching sets the value to 0.
•
For information about numeric value entry, refer to
"2.2.4 Entering numeric values/characters" (page 2-5).